Mass Flow Controllers for Gases: How Thermal MFCs Work and Where Industry Uses Them

Quick Answer
A mass flow controller measures a gas flow and holds it at a setpoint by adjusting a valve. This guide covers how thermal MFCs work, why gas calibration counts, and where industry uses them, with the Bürkert 8626 as the example.
A gas does not have a fixed volume, and that single fact causes most of the trouble in gas dosing. Ten liters per minute of nitrogen at 1 bar and ten liters per minute at 6 bar are very different amounts of nitrogen, so a rotameter that reads correctly on one day reads wrong the next when the header pressure drifts. A burner burns a mass of fuel, a bioreactor consumes a mass of oxygen, and a packaging line needs a fixed mass of carbon dioxide in every tray. For those jobs the flow has to be controlled by mass.
A mass flow controller, or MFC, is a single device that measures a gas flow by mass and holds it at a setpoint by adjusting an integrated control valve. The most common type is the thermal mass flow controller, which senses flow by how much heat the moving gas carries away from a heated element. This guide explains the mass flow controller working principle, why the calibration gas counts, how to read a datasheet honestly, where industry uses gas mass flow controllers, and what goes wrong in the field, using the Bürkert Type 8626 as the worked example.
Why Gas Flow Is Better Controlled by Mass
Volume flow of a gas depends on pressure and temperature. At constant mass, doubling the absolute pressure halves the volume, and a 30 °C rise in temperature adds about 10 percent to it. A volumetric instrument that is not compensated therefore reports a different number for the same amount of gas whenever the process conditions move.
Many processes depend on mass, not volume. Combustion needs a fixed ratio of fuel and air by mass for complete burning. A fermenter needs oxygen molecules, and the number of molecules in a liter falls as the temperature rises. Modified atmosphere packaging needs a specific blend of nitrogen and carbon dioxide, and a gas welding process needs a stable shielding gas ratio. In each case the useful quantity is the number of molecules passing per minute, which is what a mass measurement gives.
Instrument makers state mass flow in reference-condition units for this reason. A flow written as l/min (n), often called normal liters, is the volume the gas would occupy at 0 °C and 1.013 bar. A flow written in standard liters uses a reference of 20 °C or 25 °C instead, depending on the maker. The two differ by about 7 percent, so the datasheet must be read for the definition before two instruments are compared.
Mass Flow Controller Working Principle: Heat, Sensor and Valve
A thermal MFC is built from three blocks: a thermal flow sensor, a control valve, and the electronics that close the loop. The sensor does the measuring. The valve does the adjusting. The electronics compare the two and drive the valve until the measured flow matches the setpoint.
The mass flow controller working principle rests on heat transfer. A small heated element sits in or near the gas stream, with temperature sensors upstream and downstream. With no flow, the heat spreads evenly and both sensors read the same. When gas flows, it picks up heat from the element and carries it downstream, so the downstream sensor reads warmer than the upstream one. The temperature difference grows with the mass flow, because the amount of heat the gas carries depends on how many molecules pass and on their specific heat. A second method, called constant-temperature operation, holds the element at a fixed temperature above the gas and measures the electrical power needed to do so, which also rises with mass flow.
There are three physical layouts. In a capillary design, a tiny tube carries a small fraction of the flow past the heater, and a laminar flow element in the main channel keeps that fraction constant. In a MEMS design, the heater and sensors sit on a silicon chip in a similar bypass. Both suit low flows, from milliliters to a few liters per minute. An inline design puts the sensor directly in the main stream and can handle much higher flows.
The Bürkert Type 8626 belongs to the inline family. The manufacturer's data lists a thermal inline sensor in the gas stream, nominal ranges from 20 to 1500 l/min, a 1:50 turndown, and a settling time under 500 ms. That speed comes from the sensor's direct contact with the gas. A proportional control valve, driven by the electronics, adjusts the opening to hold the setpoint, and the unit communicates over an analogue signal or a fieldbus. Analogue wiring follows the usual loop conventions described in our comparison of 4-20 mA, HART and Foundation Fieldbus.
Gas Calibration and Conversion Factors
A thermal MFC measures heat transfer, and every gas carries heat differently. Helium conducts and carries heat far better than carbon dioxide does, so the same sensor output corresponds to a very different mass flow in the two gases. For that reason an MFC is calibrated for a specific gas, and the nameplate states which one.
Calibrating on every gas is expensive, so manufacturers commonly calibrate on nitrogen or air and apply a gas conversion factor to other gases. The factor is related to the ratio of the specific heats of the two gases. As an illustration, a nitrogen-calibrated sensor reading on carbon dioxide needs a factor near 0.75, so a reading of 100 units corresponds to roughly 75 units of actual CO2 flow, with the exact figure depending on the maker's table. Conversion factors work well for simple gases at moderate flow and degrade for mixtures, for gases with large molecules, and at high flow through the sensor, where the factor can drift by several percent. Brooks Instrument states the principle plainly: the highest accuracy always comes from calibrating on the actual gas.
Three consequences follow. First, if the process gas changes, the MFC needs recalibration or a new conversion setting, and changing from nitrogen to argon on a controller that has not been recalibrated produces a quiet error. Second, gas mixtures such as a nitrogen and carbon dioxide blend change their thermal properties as the ratio changes, so a controller placed after the mixing point will not read a fixed factor. The usual solution is one MFC per gas, ahead of the mixer. Third, humidity and condensation change the effective gas properties, so wet gas is a poor candidate for a thermal sensor unless it is dried first.
Reading a Datasheet Honestly: Range, Turndown, Accuracy and Response
The headline figures on a datasheet are usually correct and still misleading, because they apply under conditions the reader forgets. The table shows the Bürkert Type 8626 as published, to illustrate each term.
| Specification | Type 8626 value | What to check |
|---|---|---|
| Nominal range | 20 to 1500 l/min (n) | Which reference condition the "n" refers to |
| Turndown | 1:50 | Whether accuracy holds across the whole turndown |
| Accuracy | ±1.5 percent of reading or ±0.3 percent of full scale | How the two terms combine at low flow |
| Repeatability | ±0.1 percent of full scale | Stated at which flow |
| Settling time | Under 500 ms | The step size and the definition of settled |
| Process gas temperature | −10 to 70 °C | Whether the calibration gas matches the process |
| Supply and protection | 24 V DC, IP65 | Fieldbus or analogue option |
Take a generic controller with an accuracy stated as 1.5 percent of reading plus 0.3 percent of full scale to see why the combination counts. At 100 percent of range, the error is 1.5 percent plus 0.3 percent, or 1.8 percent of reading. At 10 percent of range, the full-scale term is 3 percent of the reading, so the total reaches 4.5 percent. At the bottom of a 1:50 range, the full-scale term alone can be 15 percent of the reading. The same instrument is accurate at the top and loose at the bottom, so the setpoint range in use decides the real accuracy, and the controller should be sized so that normal operation falls in the upper half of its range.
Datasheet accuracy also assumes a warm-up. The 8626 figure applies after a 15-minute warm-up, and a controller that is power-cycled and used immediately may sit outside its specification for a while. For a monitoring-only job, a mass flow meter for gas without a valve has the same warm-up behavior, so allow for it in any acceptance test. Response speed deserves the same scrutiny. A sub-500 ms settling time suits a burner trim loop, but a very fast loop on a long, small-bore tube can oscillate if the pipe volume downstream of the valve is large.
The sensor measures how much heat the gas carries away. It never measures how many liters went past, which is why the gas type belongs on the order and on the nameplate.
Where Industry Uses Gas Mass Flow Controllers
Combustion is the oldest use. A burner needs air and fuel gas in a controlled ratio, and a thermal mass flow controller on each stream holds that ratio as supply pressure and temperature move, which improves efficiency and reduces unburned fuel and carbon monoxide in the flue. Furnaces and heat treatment lines use protective atmospheres of nitrogen, hydrogen or blends, and an MFC on each gas sets the composition.
Food and beverage plants use mass flow control for modified atmosphere packaging, where nitrogen and carbon dioxide blends slow spoilage, for carbonation, and for nitrogen dosing in cans and kegs. Water treatment uses it too. In desalination post-treatment, carbon dioxide is dosed into permeate to restore mineralization, and ozone dosing for oxidation and disinfection needs a stable gas feed. Bioreactors and fermenters control air, oxygen and carbon dioxide feeds by mass. Laboratories and pilot plants use small MFCs for reaction gases and calibration blends. Welding shops that mix argon and carbon dioxide on site use them for shielding gas. And inerting and blanketing systems use a nitrogen MFC to keep tank pressure and oxygen level inside a safe window.
Thermal Mass Flow Meters vs Other Gas Flow Technologies
A thermal mass flow meter without a valve is the measurement half of an MFC, and it is the common choice for a mass flow meter for gas in compressed air, biogas, natural gas and flare lines. It has no moving parts, a wide turndown and almost no pressure drop. Its limits match the controller's: it depends on gas composition, it dislikes wet gas, and it cannot be recalibrated in the field without a reference.
| Technology | Measures | Strengths | Limits |
|---|---|---|---|
| Thermal mass | Mass flow directly | Wide turndown, low pressure drop, no moving parts | Gas-specific calibration, sensitive to moisture and composition |
| Coriolis | Mass flow and density | Highest accuracy, independent of composition | Higher cost, low sensitivity at very low gas density |
| Vortex | Volume flow at line conditions | Rugged, good for steam and hot gas | Needs pressure and temperature compensation for mass |
| Orifice and DP | Volume flow from pressure drop | Well established, wide acceptance | Pressure loss, narrow turndown, needs compensation |
A Coriolis meter measures mass without depending on the gas composition, and it is the benchmark where cost allows; our article on Coriolis flowmeters explains where that premium is justified. For general flowmeter choice across liquids and gases, the guide to types of industrial flowmeters covers the options, and the post on flowmeter troubleshooting lists the failure patterns that apply to thermal meters as well.
Installation and Operating Pitfalls
Most field problems with MFCs have four causes. Contamination comes first. Particles, oil mist and condensate coat the heated sensor and change its heat transfer, so the reading drifts. A filter upstream, sized for the particle size the manufacturer states, is the cheapest protection, and a coalescing filter helps where oil carry-over is possible.
Pressure comes second. An MFC needs a minimum pressure drop across its valve to control, and an inlet pressure that swings rapidly makes the loop hunt. A pressure regulator ahead of the controller stabilizes the supply. A downstream pressure that rises, for example when a packaging machine's chamber fills, can reduce the available differential until the valve cannot deliver the setpoint.
Orientation and piping come third. Many controllers are calibrated in a specific orientation, and mounting them differently shifts the zero because natural convection inside the sensor changes. Inline sensors also want a reasonably undisturbed flow profile, so a header with a valve or elbow immediately upstream gives different readings from the calibration setup.
Zero and recalibration come fourth. Thermal sensors drift slowly, so plants calibrate against a reference at intervals set by the application, commonly yearly, and more often where the gas is dirty or the accuracy requirement is tight. Controllers on toxic or flammable gases need leak-tight fittings, a purge procedure before disconnection, and, in hazardous areas, certified versions.
Real-World Scenario: A Two-Gas Packaging Blend
The figures below illustrate the arithmetic of ratio control and do not describe a specific customer.
A food plant packs fresh produce in trays under a blend of 70 percent nitrogen and 30 percent carbon dioxide by volume, at a total flow of 600 l/min (n) during filling. The target for carbon dioxide is therefore 180 l/min (n) and for nitrogen 420 l/min (n). Each gas has its own controller, ahead of a small mixing manifold, and each is calibrated on its own gas, so no conversion factor error enters.
The packaging line starts and stops with every tray, and the gas demand cycles between zero and 600 l/min. With a volumetric flow meter, the cycles would shift the mixture whenever the header pressure sagged. With mass flow controllers, each valve adjusts in under half a second, and the ratio of 420 to 180 stays fixed regardless of the pressure. A gas analyzer at the mixer outlet checks the oxygen level once per shift, which confirms that the controllers still match the process.
The plant's maintenance plan includes a filter change every quarter, a zero check at each shift start, and a reference calibration every year. It also keeps a spare controller on the shelf for the carbon dioxide line, since that line stops the packaging machine when it fails. A pair of controllers like this replaces a manual rotameter pair that needed a technician adjusting valves a dozen times a day.
Checking a Controller in the Field
A controller that is working looks the same as one that has drifted, so plants build simple checks into routine rounds. A zero check closes the valve and confirms that the flow reading falls to zero, which exposes sensor drift and leaking valves at the same time. A span check compares the controller against a reference meter in series, such as a calibrated thermal mass flow meter, at two or three setpoints across the working range, with both devices on the same gas at the same pressure.
Trend data adds a third check. If the valve opening needed to deliver a given setpoint creeps upward over weeks, the sensor is reading low, the valve seat is eroding, or the supply pressure is sagging, and any of the three justifies a visit before a quality problem appears. Controllers with a fieldbus can publish valve position and sensor temperature to the control system, so the trend costs nothing to collect. Record the zero and span results with the date, gas, pressure and serial numbers.
Choosing and Specifying a Mass Flow Controller
Selection comes down to six facts. Name the gas, and its purity and moisture content. Define the flow range in the same reference conditions as the datasheet, and place normal operation in the upper half of the range. State the inlet pressure range and the outlet pressure, so the valve has enough differential. Choose the communication: analogue 0 to 10 V or 4–20 mA, or a fieldbus for diagnostics and recipe changes, since a fieldbus lets the controller report valve position and alarms back to the PLC. Name the environment, including temperature, hazardous area class and protection rating. And decide the accuracy needed at the lowest flow actually used, not at full scale.
The Bürkert Type 8626 gas mass flow controller in the TCC catalog suits industrial gas dosing in the 20 to 1500 l/min (n) range, such as burner trim, aeration, and gas blending. Very low flows, in the milliliter range, and semiconductor or analytical duties need capillary or MEMS controllers instead, so the right device depends on the job.
Getting the gas, range and pressure facts right at the quotation stage avoids most field problems. TCC's instrumentation and control team supplies, configures and commissions mass flow controllers for industrial plants, including gas calibration checks and the supporting filters and regulators. Where the controller feeds a process recipe, our industrial automation service integrates the setpoints, alarms and totalizers into the PLC or SCADA system.
Frequently Asked Questions
What is a mass flow controller?
A mass flow controller is a device that measures the mass flow of a gas and holds it at a setpoint by adjusting an integrated control valve. It combines a flow sensor, usually thermal, a proportional valve and control electronics in one body. Industries use it wherever a gas has to be fed at a fixed mass rate, such as combustion, packaging, water treatment and fermentation.
How does a thermal mass flow controller work?
A heated element in the gas stream is flanked by temperature sensors. Flowing gas carries heat downstream, and the temperature difference, or the power needed to hold the element at a fixed temperature, is proportional to the mass flow. The electronics compare the measured flow with the setpoint and adjust the control valve until they match.
Do I need to recalibrate a mass flow controller when I change gas?
Yes, in most cases. A thermal controller is calibrated for a specific gas, and each gas carries heat differently. Manufacturers publish conversion factors for switching between gases, which work well for simple gases at moderate flow, but the best accuracy comes from calibrating on the actual process gas. Mixtures need special handling, and one controller per gas is the safer design.
What is the difference between a mass flow controller and a mass flow meter?
A mass flow meter only measures flow and reports it. A mass flow controller measures flow and also contains a control valve, so it adjusts the flow to hold a setpoint. Use a thermal mass flow meter for gas where you only need to monitor or totalize the flow, and a controller where the process needs the flow held at a set value.
Related products
Components from our catalogue relevant to this article — request a quote for availability, lead time and pricing.
/8626 Burkert.webp)
Burkert Type 8626 Mass Flow Controller for Gases (MFC)
Thermal mass flow controller for gases, 20 to 1500 l/min nominal ranges, 24 V DC, IP65
/Endress+Hauser 5P3B15-GRIRABEFAAED5S2AA1-Electromagnetic flowmeter.webp)
5P3B15-GRIRABEFAAED5S2AA1 Endress+Hauser - Electromagnetic Flowmeter
Electromagnetic flowmeter, Proline series
/Endress+Hauser 5PBB1H-AAEADBKA0AED3K0AA1-Proline Promag P 10 electromagnetic flowmeter.webp)
5PBB1H-AAEADBKA0AED3K0AA1 Endress+Hauser - Proline Promag P 10 Electromagnetic Flowmeter
Proline Promag P 10 electromagnetic flowmeter
/Endress+Hauser CLD134-PCS138AB2-Non-contact conductivity sensor.webp)
CLD134-PCS138AB2 Endress+Hauser - Non-Contact Conductivity Sensor
Inductive non-contact conductivity sensor